Neurology · 510(k) · 2025

SpineAR SNAP (SyncAR Spine)

Surgical Theater, Inc.. Cleared by FDA on 29 September 2025 under submission K252054, product code SBF (orthopedic augmented reality). What FDA's record says, in FDA's words.

Device record · K252054

  1. Device. SpineAR SNAP (SyncAR Spine)DEVICE
  2. Company. Surgical Theater, Inc.COMPANY
  3. Decision. 29 September 2025, 510(k), substantially equivalent (traditional review); received 1 July 2025DECISION
  4. Panel. NeurologyPANEL
  5. Product code. SBF, Orthopedic Augmented RealityCODE
  6. Regulation. 21 CFR 882.4560, Class IIREG
  7. Source. FDA database entry · decision document (13 pages)SOURCE

Retrieved · the fields above are FDA's; the statement below was copied from the decision document on that day.

What FDA cleared it for

The Indications for Use statement, verbatim, from the decision document.

SpineAR SNAP is intended for use for pre-operative surgical planning on-screen and in a virtual environment, and intra-operative surgical planning and visualization on-screen and in an augmented environment using the HoloLens2 AR headset display with validated navigation systems as identified in the device labeling. SpineAR SNAP is indicated for spinal stereotaxic surgery, and where reference to a rigid anatomical structure, such as the spine, can be identified relative to images of the anatomy. SpineAR is intended for use in spinal implant procedures, such as Pedicle Screw Placement, in the lumbar and thoracic regions with the HoloLens2 AR headset. The virtual display should not be relied upon solely for absolute positional information and should always be used in conjunction with the displayed 2D stereotaxic information.

K252054, FDA decision document · source

The product code's definition, from FDA's classification record. An orthopedic augmented reality device provides visual guidance during orthopedic procedures, where the use of stereotaxic navigation is utilized. Visual guidance is displayed as “augmented reality” stereoscopic images to intraoperatively augment the user’s field of view.

What the statement says

Pattern matches on the Indications for Use text. Each line is a fact that the text contains the words, not a finding about the device.

Adverse-event reports in MAUDE, by product code

As of , MAUDE held 4 reports filed against devices carrying product code SBF. That code is shared by every device of this type, not this device alone, and MAUDE reports are submitted by manufacturers, facilities and the public without FDA verification. The latest reports, in the reporter’s words:

8 July 2026 · Injury · XVISION SPINE SYSTEM (XVS) · AUGMEDICS LTD. · report 3016571711-2026-00003

AN MIS PROCEDURE WAS PERFORMED ON L4-L5 TO NAVIGATE FOUR SCREWS. FINAL FLUOROSCOPY CONFIRMATION REVEALED THAT THE SCREWS WERE POSITIONED HIGHER THAN THE PLANNED TRAJECTORY, AND THEY WERE SUBSEQUENTLY REMOVED AND SUCCESSFULLY REVISED UNDER FLUOROSCOPY. NO PATIENT HARM WAS REPORTED. THE INVESTIGATION ATTRIBUTED THE EVENT TO INTRAOPERATIVE INSTABILITY AND CONSECUTIVE SHIFTS OF THE PERC PIN REFERENCE PLATFORM, LIKELY CAUSED BY INADEQUATE SEATING WITHIN THE PSIS BONE COMBINED WITH HIGHLY UNSTABLE […]

6 May 2026 · Injury · XVISION SPINE SYSTEM (XVS) · AUGMEDICS INC. · report 3016566146-2026-00001

AN OPEN SPINAL PROCEDURE WAS PERFORMED. DURING THE PROCEDURE, NAVIGATION INACCURACIES WERE IDENTIFIED AT L2 AND S1, AND THE SCREWS WERE SUBSEQUENTLY REVISED UNDER FLUOROSCOPY. NO PATIENT HARM WAS REPORTED. THE INVESTIGATION ATTRIBUTED THE EVENT TO A PLATFORM SHIFT OF THE REFERENCE ASSEMBLY, LIKELY CAUSED BY INTRAOPERATIVE RETRACTOR MANIPULATION AND SUBOPTIMAL PIN SEATING. WHILE USER ERROR REMAINS THE MOST REASONABLE ASSUMPTION, A DEFINITIVE ROOT CAUSE COULD NOT BE CONCLUSIVELY CONFIRMED. AS […]

6 May 2026 · Injury · XVISION SPINE SYSTEM (XVS) · AUGMEDICS LTD. · report 3016571711-2026-00002

AN OPEN SPINAL PROCEDURE WAS PERFORMED. DURING THE PROCEDURE, NAVIGATION INACCURACIES WERE IDENTIFIED AT L2 AND S1, AND THE SCREWS WERE SUBSEQUENTLY REVISED UNDER FLUOROSCOPY. NO PATIENT HARM WAS REPORTED. THE INVESTIGATION ATTRIBUTED THE EVENT TO A PLATFORM SHIFT OF THE REFERENCE ASSEMBLY, LIKELY CAUSED BY INTRAOPERATIVE RETRACTOR MANIPULATION AND SUBOPTIMAL PIN SEATING. WHILE USER ERROR REMAINS THE MOST REASONABLE ASSUMPTION, A DEFINITIVE ROOT CAUSE COULD NOT BE CONCLUSIVELY CONFIRMED. AS […]

8 September 2025 · Injury · ONPOINT SURGICAL SYSTEM · ONPOINT SURGICAL, INC. · report 3016534677-2025-00001

AN MIS SURGERY ON THE LUMBAR SPINE WITH FUSION OF AND SCREW PLACEMENT IN VERTEBRAE L4 AND L5 (4 PEDICLE SCREWS TOTAL) WAS PERFORMED WITH THE AID OF THE ONPOINT SPINE SYSTEM. PATIENT WAS POSITIONED PRONE ON THE OPERATING ROOM TABLE. SURGEON ATTACHED THE NAVIGATION REFERENCE ARRAY WITH IMAGING PHANTOM TO THE PELVIS. INTRAOPERATIVE 3D CONE BEAM CT SCAN WAS PERFORMED USING AN O-ARM. SURGEON VERIFIED THE AUTOMATIC REGISTRATION OF THE SCAN TO THE PATIENT ANATOMY AND JUDGED THE ACCURACY OF THE […]

Where this fits

This page is a record, not medical advice and not a safety rating. It copies FDA's fields and FDA's statement, and lists what the statement's text mentions. Read the labeling before relying on any of it.