General and Plastic Surgery · 510(k) · 2024

HYDROS Robotic System; HYDROS TRUS Probe; HYDROS Handpiece

Procept BioRobotics. Cleared by FDA on 20 August 2024 under submission K240200, product code PZP (fluid jet removal system). What FDA's record says, in FDA's words.

Device record · K240200

  1. Device. HYDROS Robotic System; HYDROS TRUS Probe; HYDROS HandpieceDEVICE
  2. Company. Procept BioRoboticsCOMPANY
  3. Decision. 20 August 2024, 510(k), substantially equivalent (traditional review); received 25 January 2024DECISION
  4. Panel. General and Plastic SurgeryPANEL
  5. Product code. PZP, Fluid Jet Removal SystemCODE
  6. Regulation. 21 CFR 876.4350, Class IIREG
  7. Source. FDA database entry · decision document (27 pages)SOURCE

Retrieved · the fields above are FDA's; the statement below was copied from the decision document on that day.

What FDA cleared it for

The Indications for Use statement, verbatim, from the decision document.

The HYDROS Robotic System is indicated for the resection and removal of prostate tissue in males suffering from lower urinary tract symptoms (LUTS) due to benign prostatic hyperplasia. I ntended Patient Population The intended patient population is males suffering from LUTS resulting from benign prostatic hyperplasia (BPH). I ntended Users The intended user shall be a urologist, supported by OR staff, trained and familiar with performing endoscopic surgical procedures for BPH, such as TURP, and in recognizing and managing their complications. The intended user shall also be trained and familiar with TRUS imaging. T echnological Comparison as compared to the Predicate Device H YDROS ROBOTIC SYSTEM

K240200, FDA decision document · source

The product code's definition, from FDA's classification record. The system is used for removal of prostate tissue with a fluid jet.

What the statement says

Pattern matches on the Indications for Use text. Each line is a fact that the text contains the words, not a finding about the device.

Adverse-event reports in MAUDE, by product code

As of , MAUDE held 1,491 reports filed against devices carrying product code PZP. That code is shared by every device of this type, not this device alone, and MAUDE reports are submitted by manufacturers, facilities and the public without FDA verification. The latest reports, in the reporter’s words:

30 September 2026 · Injury · AQUABEAM Robotic System · PROCEPT BioRobotics Corporation · report 3012977056-2026-00199

A male patient underwent Aquablation therapy to treat symptomatic benign prostatic hyperplasia (BPH). PROCEPT BioRobotics Corporation (PROCEPT) became aware that, during Focal Bladder Neck Cautery (FBNC), the patient sustained a bladder perforation. The treating surgeon indicated that the bladder perforation likely occurred during hemostasis at a thin-walled portion of the diverticulum. The bladder was repaired laparoscopically. No malfunction of the AQUABEAM Robotic System was reported.

29 September 2026 · Malfunction · HYDROS Robotic System · PROCEPT BioRobotics Corporation · report 3012977056-2026-00198

28 September 2026 · Malfunction · AQUABEAM Robotic System · PROCEPT BioRobotics Corporation · report 3012977056-2026-00197

28 September 2026 · Injury · AQUABEAM Robotic System · PROCEPT BioRobotics Corporation · report 3012977056-2026-00196

23 September 2026 · Death · AQUABEAM Robotic System · PROCEPT BioRobotics Corporation · report 3012977056-2026-00193

A male patient underwent Aquablation therapy for symptomatic benign prostatic hyperplasia (BPH). PROCEPT BioRobotics (PROCEPT) became aware on 01-Sep-2026 that the patient had passed away. The patient experienced an acute pulmonary embolism after the first treatment pass of Aquablation therapy. The anesthesia team indicated that the patient was becoming hypotensive prior to starting the second treatment pass of Aquablation. The patient experienced cardiac arrest multiple times, and it was […]

Where this fits

This page is a record, not medical advice and not a safety rating. It copies FDA's fields and FDA's statement, and lists what the statement's text mentions. Read the labeling before relying on any of it.

More devices under product code PZP, in General and Plastic Surgery, from Procept BioRobotics, or cleared in 2024. The whole record is at FDA AI-enabled medical devices.