Radiology · 510(k) · 2020
AVIEW
Coreline Soft Co., Ltd. Cleared by FDA on 26 August 2020 under submission K200714, product code LLZ (system, image processing, radiological). What FDA's record says, in FDA's words.
Device record · K200714
- Device. AVIEWDEVICE
- Company. Coreline Soft Co., Ltd (KR)COMPANY
- Decision. 26 August 2020, 510(k), substantially equivalent (traditional review); received 18 March 2020DECISION
- Panel. RadiologyPANEL
- Product code. LLZ, System, Image Processing, RadiologicalCODE
- Regulation. 21 CFR 892.2050, Class IIREG
- Source. FDA database entry · decision document (14 pages)SOURCE
Retrieved · the fields above are FDA's; the statement below was copied from the decision document on that day.
What FDA cleared it for
The Indications for Use statement, verbatim, from the decision document.
AVIEW provides CT values for pulmonary tissue from CT thoracic and cardiac datasets. This software could be used to support the physician quantitatively in the diagnosis, follow up evaluation and documentation of CT lung tissue images by providing image segmentation of sub-structures in lung, lobe, airways and cardiac, registration of inspiration and expiration which could analyze quantitative information such as air trapping volume, air trapped index, and inspiration/ expiration ratio. And also, volumetric and structure analysis, density evaluation and reporting tools. AVIEW is also used to store, transfer, inquire and display CT data set on premise and as cloud environment as well to allow users to connect by various environment such as mobile devices and chrome browser. Characterizing nodules in the lung in a single study, or over the time course of several thoracic studies. Characterizations include nodule type, location of the nodule and measurements such as size (major axis, minor axis), estimated effective diameter from the volume of the nodule, volume of the nodule, Mean HU(the average value of the CT pixel inside the nodule in HU), Minimum HU, Max HU, mass(mass calculated from the CT pixel value), and volumetric measures(Solid major; length of the longest diameter measured in 3D for solid portion of the nodule, Solid 2nd Major: The length of the longest diameter of the solid part, measured in sections perpendicular to the Major axis of the solid portion of the […]
What the statement says
Pattern matches on the Indications for Use text. Each line is a fact that the text contains the words, not a finding about the device.
- Modalities namedCT
- Mentionsmeasurement or quantification
- Mentionsa diagnostic claim
Adverse-event reports in MAUDE, by product code
As of , MAUDE held 19,970 reports filed against devices carrying product code LLZ. That code is shared by every device of this type, not this device alone, and MAUDE reports are submitted by manufacturers, facilities and the public without FDA verification. The latest reports, in the reporter’s words:
30 September 2026 · Malfunction · IntelliSpace Cardiovascular · Philips Medical Systems Nederland B.V. · report 3003768277-2026-101631
Philips received a complaint on the IntelliSpace Cardiovascular (ISCV) indicating that the ISCV could only be accessed in Web-Only mode on the workstation. The device was in clinical use at the time of the event, and no adverse events or harm were reported in the complaint ((b)(4)).Philips has started an investigation of this complaint.
30 September 2026 · Malfunction · IntelliSpace Cardiovascular · Philips Medical Systems Nederland B.V. · report 3003768277-2026-101632
Philips received a complaint on the IntelliSearch Cardiovascular (ISCV) indicating that customer reported that study retrieval from the archive was either failing or taking an excessively long time to complete. The device was in clinical use at the time of the event, and no adverse events or harm were reported in the complaint (b)(4).Philips has started an investigation of this complaint.
30 September 2026 · Malfunction · IntelliSpace CardioVascular · Philips Medical Systems Nederland B.V. · report 3003768277-2026-101634
Philips identified a software defect in ISCV (IntelliSpace Cardiovascular) system wherein the user was reported that DICOM import crashed occasionally as there was a PDFClinicalContentExecuter error.. The device was in clinical use at the time of event.Based on the new information that Philips became aware, this has been determined to be a reportable malfunction. Under specific circumstances, it could potentially lead to delayed diagnosis. Given the defect¿s nature and its potential impact on […]
30 September 2026 · Malfunction · IntelliSpace CardioVascular · Philips Medical Systems Nederland B.V. · report 3003768277-2026-101635
Philips identified a software defect in ISCV (IntelliSpace Cardiovascular) system wherein the user could not save the findings. The device was in clinical use at the time of event. Based on the new information that Philips became aware, this has been determined to be a reportable malfunction. Under specific circumstances, it could potentially lead to delayed diagnosis. Given the defect¿s nature and its potential impact on clinical decision-making if it were to recur, it is determined that this […]
30 September 2026 · Malfunction · IntelliSpace Cardiovascular · Philips Medical Systems Nederland B.V. · report 3003768277-2026-101636
Philips identified a software defect in ISCV (IntelliSpace Cardiovascular) system wherein user reported that it was not possible to view previous studies in Echo Module. The device was in clinical use at the time of event. Based on the new information that Philips became aware, this has been determined to be a reportable malfunction. Under specific circumstances, it could potentially lead to delayed diagnosis. Given the defect¿s nature and its potential impact on clinical decision-making if it […]
Where this fits